 |
| Name: |
HP 5890 Series
II with FPD |
| Item: |
#589 |
| Pirce: |
|
| |
|
| The HP 5890 II GC is in good
condition and refurbished at GenTech. Includes
single Split/Splitless Injector, single Packed
Injector, single Flame Photometric Detector
(FPD), INET Communications installed on GC,
cables, manuals and NINETY
(90) DAY WARRANTY. HPIB Communications
and Data System available. |
 |
|
Back
|
|
| Specifications: |
| Heated Devices |
Five heated zones standard: Two detectors, Two inlets,
one auxiliary
Methods stored: Two |
| Split/Splitless
Capillary |
Range to 400°C in 1°C increments
Back-pressure design permits independent adjustment
of split flow rate without affecting column flow
Septum purge built in at 3 ml/min
Accepts columns up to 1.2-mm od
Multimode design includes split and splitless injection
Splitless purge time variable in 0.01-min elements |
| Flame Ionization
Detector |
Range to 450°C
Grounded jet and current limited design for operator
safety
Push-button flame ignition
Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in.
id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in.
id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in.
id capillary jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in.
id capillary jet
Minimum detectable: <5 pg carbon/sec, nitrogen
carrier at S/N=2
Linear dynamic range: <+/- 10% over a 107
range with 0.018-in. id jet
Conditions: column flow 50 ml/min, 45 ml/min H2,
650 ml/min air, propane sample |