GC
Agilent / HP Gas Chromatograph Systems
Name: HP 5890 Series II with EPC, Dual FID
Item: #018
Pirce:  
 
Includes dual split/splitless injectors with Electronic Pressure Control, dual Flame Ionization Detectors, INET Communications installed on GC, cables, manuals, and ONE (1) YEAR WARRANTY. Other injectors and detectors available. HPIB Communications and Data System available.
Back

Specifications:
Heated Devices
Five heated zones standard: Two detectors, Two inlets, one auxiliary
Methods stored: Two
Split/Splitless Capillary
Range to 400°C in 1°C increments
Back-pressure design permits independent adjustment of split flow rate without affecting column flow
Septum purge built in at 3 ml/min
Accepts columns up to 1.2-mm od
Multimode design includes split and splitless injection
Splitless purge time variable in 0.01-min elements
Electronic Pressure Control
Pressure Control
> Programming: Three ramps, 0.1-99 psi/minute
> Pressure Range for inlets: 1-100 psi
> Flow range for inlets: 0-450 ml/min
> Constant Flow Mode (Standard): Inlet pressure is adjusted real time to maintain constant volumetric flow at column outlet.
> Maximum Channels: Two inlet channels to control column head pressure, Four auxiliary channels to control split vent flow (Gas Saver EPC), makeup gas, fuel gases, or external devices (general-purpose EPC)
Precision and Accuracy
> Pressure Setpoint Accuracy: +/- 0.1 psi
> Flow Setpoint Accuracy: +/- 0.001 ml/min
> Average Linear Velocity Accuracy: +/- 0.1 0.001 cm/sec.
Flame Ionization Detector
Range to 450°C
Grounded jet and current limited design for operator safety
Push-button flame ignition
Sensitive:
> 18 mCoul/gm carbon: nitrogen carrier, 0.018-in. id jet
> 15 mCoul/gm carbon: helium carrier, 0.018-in. id jet
> 22mCoul/gm carbon: nitrogen carrier, 0.018-in. id capillary jet
> 18 mCoul/gm carbon: helium carrier, 0.018-in. id capillary jet
Minimum detectable: <5 pg carbon/sec, nitrogen carrier at S/N=2
Linear dynamic range: <+/- 10% over a 107 range with 0.018-in. id jet
Conditions: column flow 50 ml/min, 45 ml/min H2, 650 ml/min air, propane sample
Column Oven
Operating range: 4° C above ambient to 450° C
Setpoint entry: 1° C for temperatures, 0.1° C for program rates
Programming: Rates 0.1 to 70° C per min, 650 min maximum run time, three ramps with initial/final holds